Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading
BACKGROUND: Rigorous efforts to reduce the recurrence of abutment screw loosening in single-tooth implant restorations have recently been made. However, the behavior of the implant/abutment joint components with respect to critical bending force is still unclear.
PURPOSE: This study investigated the effect of different cyclic loading periods on abutment screw loosening and bending resistance of a single-tooth external hexagon implant system.
MATERIAL AND METHODS: Fifteen Brånemark implant assemblies were divided equally into groups A, B, and C. Each assembly consisted of a Brånemark System Mk IV 4 x 10 mm implant (Nobel Biocare AB, Gothenburg, Sweden) mounted in a brass block, a CeraOne 3 mm abutment (Nobel Biocare AB), and an experimental cement-retained superstructure. For groups A and B a cyclic load of 50 N was applied centrally and perpendicular to the long axis of the implant. Targets of 1.0 x 10(6) cycles (40 months of simulated function) and 0.5 x 10(6) cycles (20 months of simulated function) were defined for groups A and B, respectively. Group C (control) was left unloaded for the same loading time period as was group B. Reverse torque was recorded before and after loading, and the difference was calculated. After cyclic loading, specimens were mounted in a testing machine, and the yielding and bending strengths were measured. The data were analyzed with one-way analysis of variance and were compared by means of the Tukey test (p < .05).
RESULTS: There were statistically significant differences (p < .001) in the reverse torque difference values of group A ([-5.6 to -3.4] +/- 0.86 Ncm) as compared to those of group B ([-2.4 to -1.6] +/- 0.32 Ncm) and group C ([-0.7 to 0.0] +/- 0.26 Ncm). Likewise, group B showed a significant difference compared to group C (p = .002). On the other hand there was no statistically significant difference in the mean values among the test groups in regard to the yielding and bending strengths (p >.050).
CONCLUSION: Within the limitations of this study, long-term fatigue significantly affected the reverse torque values under centric lateral load (p <.001) whereas it had no significant effect on the resistance of the implant/abutment joint to static bending.
Medienart: |
Artikel |
---|
Erscheinungsjahr: |
2004 |
---|---|
Erschienen: |
2004 |
Enthalten in: |
Zur Gesamtaufnahme - volume:6 |
---|---|
Enthalten in: |
Clinical implant dentistry and related research - 6(2004), 3 vom: 01., Seite 157-64 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Khraisat, Ameen [VerfasserIn] |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 08.03.2005 Date Revised 09.11.2019 published: Print Citation Status MEDLINE |
---|
Förderinstitution / Projekttitel: |
|
---|
PPN (Katalog-ID): |
NLM153828951 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM153828951 | ||
003 | DE-627 | ||
005 | 20231223064809.0 | ||
007 | tu | ||
008 | 231223s2004 xx ||||| 00| ||eng c | ||
028 | 5 | 2 | |a pubmed24n0513.xml |
035 | |a (DE-627)NLM153828951 | ||
035 | |a (NLM)15726850 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Khraisat, Ameen |e verfasserin |4 aut | |
245 | 1 | 0 | |a Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading |
264 | 1 | |c 2004 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ohne Hilfsmittel zu benutzen |b n |2 rdamedia | ||
338 | |a Band |b nc |2 rdacarrier | ||
500 | |a Date Completed 08.03.2005 | ||
500 | |a Date Revised 09.11.2019 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a BACKGROUND: Rigorous efforts to reduce the recurrence of abutment screw loosening in single-tooth implant restorations have recently been made. However, the behavior of the implant/abutment joint components with respect to critical bending force is still unclear | ||
520 | |a PURPOSE: This study investigated the effect of different cyclic loading periods on abutment screw loosening and bending resistance of a single-tooth external hexagon implant system | ||
520 | |a MATERIAL AND METHODS: Fifteen Brånemark implant assemblies were divided equally into groups A, B, and C. Each assembly consisted of a Brånemark System Mk IV 4 x 10 mm implant (Nobel Biocare AB, Gothenburg, Sweden) mounted in a brass block, a CeraOne 3 mm abutment (Nobel Biocare AB), and an experimental cement-retained superstructure. For groups A and B a cyclic load of 50 N was applied centrally and perpendicular to the long axis of the implant. Targets of 1.0 x 10(6) cycles (40 months of simulated function) and 0.5 x 10(6) cycles (20 months of simulated function) were defined for groups A and B, respectively. Group C (control) was left unloaded for the same loading time period as was group B. Reverse torque was recorded before and after loading, and the difference was calculated. After cyclic loading, specimens were mounted in a testing machine, and the yielding and bending strengths were measured. The data were analyzed with one-way analysis of variance and were compared by means of the Tukey test (p < .05) | ||
520 | |a RESULTS: There were statistically significant differences (p < .001) in the reverse torque difference values of group A ([-5.6 to -3.4] +/- 0.86 Ncm) as compared to those of group B ([-2.4 to -1.6] +/- 0.32 Ncm) and group C ([-0.7 to 0.0] +/- 0.26 Ncm). Likewise, group B showed a significant difference compared to group C (p = .002). On the other hand there was no statistically significant difference in the mean values among the test groups in regard to the yielding and bending strengths (p >.050) | ||
520 | |a CONCLUSION: Within the limitations of this study, long-term fatigue significantly affected the reverse torque values under centric lateral load (p <.001) whereas it had no significant effect on the resistance of the implant/abutment joint to static bending | ||
650 | 4 | |a Journal Article | |
700 | 1 | |a Abu-Hammad, Osama |e verfasserin |4 aut | |
700 | 1 | |a Dar-Odeh, Najla |e verfasserin |4 aut | |
700 | 1 | |a Al-Kayed, Awni M |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Clinical implant dentistry and related research |d 1999 |g 6(2004), 3 vom: 01., Seite 157-64 |w (DE-627)NLM112652921 |x 1708-8208 |7 nnns |
773 | 1 | 8 | |g volume:6 |g year:2004 |g number:3 |g day:01 |g pages:157-64 |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 6 |j 2004 |e 3 |b 01 |h 157-64 |