Time-limited efficacy of pacing electrodes following open heart surgery
Temporary epicardial pacing electrodes have been utilised since the 1960s in the postoperative management of cardiac surgical patients, both as a diagnostic tool and therapeutic intervention. To determine the efficacy of the epicardial pacing wires over time after open heart surgery, 30 patients (20M/10F) who underwent coronary artery bypass surgery, were evaluated by standard 12-lead EKG, atrial electrogram, and atrial and ventricular pacing thresholds immediately after surgery and on postoperative day 5. Both atrial and ventricular pacing thresholds were significantly increased on postoperative day 5 as compared to baseline. The ability for effective AAI, VVI and DDD pacing was lost in 38.89 percent, 37.5 percent and 61.11 percent of patients, respectively, on postoperative day 5. We conclude that both atrial and ventricular pacing wires have limited efficacy after postoperative day 4 for pacing after open heart surgery due to a marked increase in pacing thresholds over this time period.
Medienart: |
Artikel |
---|
Erscheinungsjahr: |
1996 |
---|---|
Erschienen: |
1996 |
Enthalten in: |
Zur Gesamtaufnahme - volume:48 |
---|---|
Enthalten in: |
Indian heart journal - 48(1996), 6 vom: 22. Nov., Seite 681-4 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Kosmas, C E [VerfasserIn] |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 10.04.1997 Date Revised 15.11.2007 published: Print Citation Status MEDLINE |
---|
Förderinstitution / Projekttitel: |
|
---|
PPN (Katalog-ID): |
NLM090243951 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM090243951 | ||
003 | DE-627 | ||
005 | 20231222082115.0 | ||
007 | tu | ||
008 | 231222s1996 xx ||||| 00| ||eng c | ||
028 | 5 | 2 | |a pubmed24n0301.xml |
035 | |a (DE-627)NLM090243951 | ||
035 | |a (NLM)9062018 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Kosmas, C E |e verfasserin |4 aut | |
245 | 1 | 0 | |a Time-limited efficacy of pacing electrodes following open heart surgery |
264 | 1 | |c 1996 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ohne Hilfsmittel zu benutzen |b n |2 rdamedia | ||
338 | |a Band |b nc |2 rdacarrier | ||
500 | |a Date Completed 10.04.1997 | ||
500 | |a Date Revised 15.11.2007 | ||
500 | |a published: Print | ||
500 | |a Citation Status MEDLINE | ||
520 | |a Temporary epicardial pacing electrodes have been utilised since the 1960s in the postoperative management of cardiac surgical patients, both as a diagnostic tool and therapeutic intervention. To determine the efficacy of the epicardial pacing wires over time after open heart surgery, 30 patients (20M/10F) who underwent coronary artery bypass surgery, were evaluated by standard 12-lead EKG, atrial electrogram, and atrial and ventricular pacing thresholds immediately after surgery and on postoperative day 5. Both atrial and ventricular pacing thresholds were significantly increased on postoperative day 5 as compared to baseline. The ability for effective AAI, VVI and DDD pacing was lost in 38.89 percent, 37.5 percent and 61.11 percent of patients, respectively, on postoperative day 5. We conclude that both atrial and ventricular pacing wires have limited efficacy after postoperative day 4 for pacing after open heart surgery due to a marked increase in pacing thresholds over this time period | ||
650 | 4 | |a Journal Article | |
700 | 1 | |a Ryder, R G |e verfasserin |4 aut | |
700 | 1 | |a Poon, M J |e verfasserin |4 aut | |
700 | 1 | |a Zelenkofske, S L |e verfasserin |4 aut | |
700 | 1 | |a Banka, V S |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Indian heart journal |d 1961 |g 48(1996), 6 vom: 22. Nov., Seite 681-4 |w (DE-627)NLM000004162 |x 2213-3763 |7 nnns |
773 | 1 | 8 | |g volume:48 |g year:1996 |g number:6 |g day:22 |g month:11 |g pages:681-4 |