Optimal Strategy for HIFU-Based Renal Sympathetic Denervation in Canines

Background: The association between the treatment efficacy and safety of high-intensity focused ultrasound (HIFU)-based renal sympathetic denervation (RDN) and the acoustic energy dose applied has not been fully studied and may provide important understanding of the mechanism that led to failure of the WAVE IV trial. The objective of this study was to externally deliver different HIFU doses to canines for RDN treatment and to investigate the optimal energy dose for HIFU-based RDN.Methods: Thirty canines were divided into five RDN groups according to dose of acoustic energy applied, and a sham control group that consisted of four canines was used for comparisons. All animals in the RDN groups underwent the RDN procedure with different acoustic energy doses, while in the sham control group, renal arteries were harvested without being subjected to acoustic energy delivery and were imaged using color Doppler flow imaging (CDFI). Blood pressure (BP) was recorded, and blood samples were collected before the RDN procedure and at 28 days after the RDN procedure. Histological examinations and measurement of renal tissue norepinephrine concentration were performed in all retrieved samples.Results: Suppression of BP was significant in the 300 W (15.17/8.33 ± 1.47/1.21 mmHg), 250 W (14.67/9.33 ± 1.21/1.37 mmHg), and 200 W (13.17/9.17 ± 2.32/1.84 mmHg) groups. Semiquantitative histological assessment of periarterial nerves around the kidney revealed that target nerves in the 300 W (9.77 ± 0.63), 250 W (9.42 ± 0.67), and 200 W (9.58 ± 0.54) groups had the highest nerve injury scores, followed by the 150 W group (5.29 ± 0.62). Furthermore, decreased renal tissue norepinephrine concentration, together with decreased expression of tyrosine hydroxylase in the 300, 250, and 200 W groups demonstrated effective sympathetic depression following sufficient acoustic energy deposition. However, the renal artery injury score in the 300 W group (0.93 ± 0.13) was significantly higher than in the other groups (p < 0.001).Conclusion: This study provides evidence that RDN effectiveness is based on the energy dose delivered and that 200–250 W is effective and safe in normal-sized canines..

Medienart:

E-Artikel

Erscheinungsjahr:

2021

Erschienen:

2021

Enthalten in:

Zur Gesamtaufnahme - volume:8

Enthalten in:

Frontiers in Cardiovascular Medicine - 8(2021)

Sprache:

Englisch

Beteiligte Personen:

Qingyao Liao [VerfasserIn]
Fang Yang [VerfasserIn]
Bo Xiong [VerfasserIn]
Xiaoyu Zheng [VerfasserIn]
Yue Wang [VerfasserIn]
Jun Qian [VerfasserIn]
Zhao Qiao [VerfasserIn]
Jing Huang [VerfasserIn]

Links:

doi.org [kostenfrei]
doaj.org [kostenfrei]
www.frontiersin.org [kostenfrei]
Journal toc [kostenfrei]

Themen:

Diseases of the circulatory (Cardiovascular) system
High intensity focused ultrasound
Hypertension
Pre-clinical study
Renal sympathetic denervation
Ultrasound

doi:

10.3389/fcvm.2021.739560

funding:

Förderinstitution / Projekttitel:

PPN (Katalog-ID):

DOAJ054777801