Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse oximeters and electrocardiography
OBJECTIVE: To compare the efficacy and reliability of two pulse-oximeters (POx) (Masimo Radical-7 and Nellcor™ Oxymax Bedside) and evaluate the feasibility of routine ECG monitoring during delivery room transition.
STUDY DESIGN: Prospective observational comparative study. Sixty newborns were connected simultaneously to both POxs and ECG monitor (as a gold standard for HR). Times to achieve a stable signal were compared. Heart rates were compared to simultaneous ECG.
RESULTS: A significant difference in times to stable signal was found: Mean, Median (Interquartile range) for Nellcor and Masimo, were 15, 8.5 (6-18) and 27, 12 (9-34) seconds, respectively. Compared to ECG, false bradycardia was displayed in 18 of 55 (35%) newborns by the Masimo POx and in no newborns by the Nellcor POx. Attaching the ECG monitor was feasible but consumed additional resources.
CONCLUSIONS: The time for achievement of a stable saturation reading in an uncomplicated resuscitation setting differed significantly between POxs.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2021 |
---|---|
Erschienen: |
2021 |
Enthalten in: |
Zur Gesamtaufnahme - volume:41 |
---|---|
Enthalten in: |
Journal of perinatology : official journal of the California Perinatal Association - 41(2021), 4 vom: 30. Apr., Seite 885-890 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Khoury, Rasha [VerfasserIn] |
---|
Links: |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 31.08.2021 Date Revised 01.09.2021 published: Print-Electronic Citation Status MEDLINE |
---|
doi: |
10.1038/s41372-020-00881-y |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM318169088 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM318169088 | ||
003 | DE-627 | ||
005 | 20231225164807.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2021 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1038/s41372-020-00881-y |2 doi | |
028 | 5 | 2 | |a pubmed24n1060.xml |
035 | |a (DE-627)NLM318169088 | ||
035 | |a (NLM)33250516 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Khoury, Rasha |e verfasserin |4 aut | |
245 | 1 | 0 | |a Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse oximeters and electrocardiography |
264 | 1 | |c 2021 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 31.08.2021 | ||
500 | |a Date Revised 01.09.2021 | ||
500 | |a published: Print-Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a OBJECTIVE: To compare the efficacy and reliability of two pulse-oximeters (POx) (Masimo Radical-7 and Nellcor™ Oxymax Bedside) and evaluate the feasibility of routine ECG monitoring during delivery room transition | ||
520 | |a STUDY DESIGN: Prospective observational comparative study. Sixty newborns were connected simultaneously to both POxs and ECG monitor (as a gold standard for HR). Times to achieve a stable signal were compared. Heart rates were compared to simultaneous ECG | ||
520 | |a RESULTS: A significant difference in times to stable signal was found: Mean, Median (Interquartile range) for Nellcor and Masimo, were 15, 8.5 (6-18) and 27, 12 (9-34) seconds, respectively. Compared to ECG, false bradycardia was displayed in 18 of 55 (35%) newborns by the Masimo POx and in no newborns by the Nellcor POx. Attaching the ECG monitor was feasible but consumed additional resources | ||
520 | |a CONCLUSIONS: The time for achievement of a stable saturation reading in an uncomplicated resuscitation setting differed significantly between POxs | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Observational Study | |
650 | 7 | |a Oxygen |2 NLM | |
650 | 7 | |a S88TT14065 |2 NLM | |
700 | 1 | |a Klinger, Gil |e verfasserin |4 aut | |
700 | 1 | |a Shir, Yakir |e verfasserin |4 aut | |
700 | 1 | |a Osovsky, Michael |e verfasserin |4 aut | |
700 | 1 | |a Bromiker, Ruben |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Journal of perinatology : official journal of the California Perinatal Association |d 1993 |g 41(2021), 4 vom: 30. Apr., Seite 885-890 |w (DE-627)NLM01262943X |x 1476-5543 |7 nnns |
773 | 1 | 8 | |g volume:41 |g year:2021 |g number:4 |g day:30 |g month:04 |g pages:885-890 |
856 | 4 | 0 | |u http://dx.doi.org/10.1038/s41372-020-00881-y |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 41 |j 2021 |e 4 |b 30 |c 04 |h 885-890 |