Monitoring lung impedance changes during long-term ventilator-induced lung injury ventilation using electrical impedance tomography
OBJECTIVE: The target of this methodological evaluation was the feasibility of long-term monitoring of changes in lung conditions by time-difference electrical impedance tomography (tdEIT). In contrast to ventilation monitoring by tdEIT, the monitoring of end-expiratory (EELIC) or end-inspiratory (EILIC) lung impedance change always requires a reference measurement.
APPROACH: To determine the stability of the used Pulmovista 500® EIT system, as a prerequisite it was initially secured on a resistive phantom for 50 h. By comparing the slopes of EELIC for the whole lung area up to 48 h from 36 pigs ventilated at six positive end-expiratory pressure (PEEP) levels from 0 to 18 cmH2O we found a good agreement (range of r 2 = 0.93-1.0) between absolute EIT (aEIT) and tdEIT values. This justified the usage of tdEIT with its superior local resolution compared to aEIT for long-term determination of EELIC.
MAIN RESULTS: The EELIC was between -0.07 Ωm day-1 at PEEP 4 and -1.04 Ωm day-1 at PEEP 18 cmH2O. The complex local time pattern for EELIC was roughly quantified by the new parameter, centre of end-expiratory change (CoEEC), in equivalence to the established centre of ventilation (CoV). The ventrally located mean of the CoV was fairly constant in the range of 42%-46% of thorax diameter; however, on the contrary, the CoEEC shifted from about 40% to about 75% in the dorsal direction for PEEP levels of 14 and 18 cmH2O.
SIGNIFICANCE: The observed shifts started earlier for higher PEEP levels. Changes of EELI could be precisely monitored over a period of 48 h by tdEIT on pigs.
Medienart: |
E-Artikel |
---|
Erscheinungsjahr: |
2020 |
---|---|
Erschienen: |
2020 |
Enthalten in: |
Zur Gesamtaufnahme - volume:41 |
---|---|
Enthalten in: |
Physiological measurement - 41(2020), 9 vom: 09. Okt., Seite 095011 |
Sprache: |
Englisch |
---|
Beteiligte Personen: |
Hahn, G [VerfasserIn] |
---|
Links: |
---|
Themen: |
---|
Anmerkungen: |
Date Completed 23.10.2020 Date Revised 23.10.2020 published: Electronic Citation Status MEDLINE |
---|
doi: |
10.1088/1361-6579/abb1fb |
---|
funding: |
|
---|---|
Förderinstitution / Projekttitel: |
|
PPN (Katalog-ID): |
NLM316051527 |
---|
LEADER | 01000naa a22002652 4500 | ||
---|---|---|---|
001 | NLM316051527 | ||
003 | DE-627 | ||
005 | 20231225160237.0 | ||
007 | cr uuu---uuuuu | ||
008 | 231225s2020 xx |||||o 00| ||eng c | ||
024 | 7 | |a 10.1088/1361-6579/abb1fb |2 doi | |
028 | 5 | 2 | |a pubmed24n1053.xml |
035 | |a (DE-627)NLM316051527 | ||
035 | |a (NLM)33035199 | ||
040 | |a DE-627 |b ger |c DE-627 |e rakwb | ||
041 | |a eng | ||
100 | 1 | |a Hahn, G |e verfasserin |4 aut | |
245 | 1 | 0 | |a Monitoring lung impedance changes during long-term ventilator-induced lung injury ventilation using electrical impedance tomography |
264 | 1 | |c 2020 | |
336 | |a Text |b txt |2 rdacontent | ||
337 | |a ƒaComputermedien |b c |2 rdamedia | ||
338 | |a ƒa Online-Ressource |b cr |2 rdacarrier | ||
500 | |a Date Completed 23.10.2020 | ||
500 | |a Date Revised 23.10.2020 | ||
500 | |a published: Electronic | ||
500 | |a Citation Status MEDLINE | ||
520 | |a OBJECTIVE: The target of this methodological evaluation was the feasibility of long-term monitoring of changes in lung conditions by time-difference electrical impedance tomography (tdEIT). In contrast to ventilation monitoring by tdEIT, the monitoring of end-expiratory (EELIC) or end-inspiratory (EILIC) lung impedance change always requires a reference measurement | ||
520 | |a APPROACH: To determine the stability of the used Pulmovista 500® EIT system, as a prerequisite it was initially secured on a resistive phantom for 50 h. By comparing the slopes of EELIC for the whole lung area up to 48 h from 36 pigs ventilated at six positive end-expiratory pressure (PEEP) levels from 0 to 18 cmH2O we found a good agreement (range of r 2 = 0.93-1.0) between absolute EIT (aEIT) and tdEIT values. This justified the usage of tdEIT with its superior local resolution compared to aEIT for long-term determination of EELIC | ||
520 | |a MAIN RESULTS: The EELIC was between -0.07 Ωm day-1 at PEEP 4 and -1.04 Ωm day-1 at PEEP 18 cmH2O. The complex local time pattern for EELIC was roughly quantified by the new parameter, centre of end-expiratory change (CoEEC), in equivalence to the established centre of ventilation (CoV). The ventrally located mean of the CoV was fairly constant in the range of 42%-46% of thorax diameter; however, on the contrary, the CoEEC shifted from about 40% to about 75% in the dorsal direction for PEEP levels of 14 and 18 cmH2O | ||
520 | |a SIGNIFICANCE: The observed shifts started earlier for higher PEEP levels. Changes of EELI could be precisely monitored over a period of 48 h by tdEIT on pigs | ||
650 | 4 | |a Journal Article | |
650 | 4 | |a Research Support, Non-U.S. Gov't | |
700 | 1 | |a Niewenhuys, J |e verfasserin |4 aut | |
700 | 1 | |a Just, A |e verfasserin |4 aut | |
700 | 1 | |a Tonetti, T |e verfasserin |4 aut | |
700 | 1 | |a Behnemann, T |e verfasserin |4 aut | |
700 | 1 | |a Rapetti, F |e verfasserin |4 aut | |
700 | 1 | |a Collino, F |e verfasserin |4 aut | |
700 | 1 | |a Vasques, F |e verfasserin |4 aut | |
700 | 1 | |a Maiolo, G |e verfasserin |4 aut | |
700 | 1 | |a Romitti, F |e verfasserin |4 aut | |
700 | 1 | |a Gattinoni, L |e verfasserin |4 aut | |
700 | 1 | |a Quintel, M |e verfasserin |4 aut | |
700 | 1 | |a Moerer, O |e verfasserin |4 aut | |
773 | 0 | 8 | |i Enthalten in |t Physiological measurement |d 1997 |g 41(2020), 9 vom: 09. Okt., Seite 095011 |w (DE-627)NLM074655876 |x 1361-6579 |7 nnns |
773 | 1 | 8 | |g volume:41 |g year:2020 |g number:9 |g day:09 |g month:10 |g pages:095011 |
856 | 4 | 0 | |u http://dx.doi.org/10.1088/1361-6579/abb1fb |3 Volltext |
912 | |a GBV_USEFLAG_A | ||
912 | |a GBV_NLM | ||
951 | |a AR | ||
952 | |d 41 |j 2020 |e 9 |b 09 |c 10 |h 095011 |